Features:
- Benzene specific measurement
- 0.01 – 50 ppm with 10.0 eV PID lamp and pre-filter tube
- Benzene select mode takes timed samples using pre-filter tube
- Long break through times for benzene filter tubes
- 10.0 eV Selective PID sensor
- 0 – 100 ppm VOC range
- Library of over 300 VOC’s including BTEX
- Humidity resistant
- Easily switched VOC target gas
- Favorite list of 30 VOC’s
- Recently used list of 8 VOC’s
- Dual channel PIDs utilizing both 10.0 & 10.6 PID sensors
- 2 smart sensor slots auto recognize PID, IR, and super toxic sensors
- Internal sample pump
- Dust & Water resistant IP-67 design
- Interchangeable battery packs Li-ion / alkaline
- Operates up to 14 hours on Li-ion battery pack, fully recharges in 3 hours
Application/s:
-
Petroleum / Oil refining
-
Chemical manufacturing
-
Rubber & Tire manufacturing
-
Coke & Coal manufacturing
-
Fuel Storage & Tank Farms
-
Hazmat response teams
Overview of the Benzene Personal PID gas monitor:
The GX-6000 personal Benzene PID gas monitor is the lightest and smallest Benzene monitor created, weighing in at only 600 grams with the rubber boot.
Simply put, the GX-6000 is a game changer with Benzene Specific measuring mode. With a library of over 300 VOC’s and the addition of the highly selective 10.0 eV PID sensor, the GX-6000 can now detect the BTEX (benzene, toluene, ethylbenzene, and xylene) family of VOC’s up to 100 ppm. By using filtering tubes and the tube holder, the GX-6000 can monitor specifically for Benzene at a range of 0 – 50ppm (with resolutions as small as 0.01) by filtering out all other VOC’s. The filter tube holder has a unique break away design for accidental damage. The filter tube also has a safe and convenient glass breaker for the filter tube tips.
The GX-6000 is the only PID instrument in the market with the ability to utilise two PID sensors in one instrument. When using the broad range 10.6eV PID for preliminary investigation, it is possible to switch to benzene mode for a highly accurate measurement.
The GX-6000 has a rugged design built for the nastiest environments. It is equipped with a removable impact-resistant rubber boot and a dust and water resistant enclosure with an IP-67 rating. With 5 bright LED lights on 3 sides of the instrument, alarms are easily seen from a variety of perspectives. With vibration and loud audible alarm, GX-6000 will easily alert users when needed.
Specifications:
Alarm Types | Gas alarm | Man-down | Trouble |
Low and high adjustable | 1st – self reset
2nd – latching |
Low flow
Sensor connection Low battery Circuit error Calibration range
|
|
Sample Method | Sample draw pump, flow rate 0.45 LPM, 30m sample range | ||
Display | Digital LCD, autobacklite, peak bar display, auto display rotation, and customize order of gases | ||
Audible Alarm | 95 dB at 1 ft. | ||
Languages | English, German, Japanese, Korean, Russian | ||
Continuous Operation * |
14 hours on Lithium-ion battery pack. Complete recharge in 3 hours.
8 hours Alkaline battery pack (3 AA size within each pack) |
||
Operating Temp. & Humidity |
-4°F to 122°F (-20°C ~ +50?), 0- 95% RH non condensing | ||
Dimensions / Weight | 2.6” W x 7.8” H x 2” D (H200 x W68 x D52 mm) 14 ounce (400g) (with lithium battery pack) |
||
Case Material | Dust and water resistant with IP-67 approval (exclude sample hose and probe) | ||
Safety Design /
Approvals |
Explosion proof: IECEx , ATEX | ||
Standard |
|
||
Standard Accessories |
|
||
Optional Accessories |
|
||
Warranty | Two years material (including sensors) and workmanship. One year for PID sensor |
Gas | Measuring Range | Minimum Display | Alarm Set Points | Accuracy * Which ever is greater |
PID Sensors | ||||
10.6 PID Lamp 600 VOC’s |
0 – 50,000 ppb | 1 ppb (< 5000 ppb) 10 ppb (> 5,000 ppb) |
A1: 5,000 ppb | +5% of full scale |
A2: 10,000 ppb | ||||
0 – 6,000 ppm | 0.1 ppm (< 600 ppm) 1 ppm (> 600 ppm) |
A1: 400 ppm | ||
A2: 1,000 ppm | ||||
10.0 PID Lamp
Benzene Specific No Alarms |
Benzene Specific 0.1 – 50 ppm |
0.01 ppm (0 ~ 10 ppm) 0.1 ppm (10 ~ 50 ppm) |
no alarms | |
300 VOC’s 0 – 100 ppm |
0.01 ppm (0 ~ 10 ppm) 0.1 ppm (10 ~ 100 ppm) |
A1: 5 ppm | ||
A2: 10 ppm | ||||
Catalytic | ||||
Hydrocarbons (CH4, std) | 0 – 100% LEL | 1% LEL | A1: 10%LEL | ± 5% of reading or ± 2% LEL (*) |
A2: 50%LEL | ||||
Galvanic | ||||
Oxygen (O2) | 0 – 40% Vol. | 0.1% Vol. | A1: 19.5 Vol. | ± 0.5% O2 |
A2: 23.5 Vol. | ||||
Electro Chemical | ||||
Carbon Monoxide (CO) | 0 – 500 ppm | 1 ppm | A1: 25 ppm | ± 5% of reading or ± 5 ppm |
A2: 50 ppm | ||||
TWA 25 ppm | ||||
STEL 200 ppm | ||||
Hydrogen Sulfide (H2S) | 0 – 100 ppm | 0.5 ppm | A1: 5.0 ppm | ± 5% of reading or± 2 ppm |
A2: 30.0 ppm | ||||
TWA 1.0 ppm | ||||
STEL 5.0 ppm | ||||
Toxics (EC Sensors) | ||||
Ammonia (NH3)* | 0 – 75 ppm
0 – 400 ppm |
0.5 ppm
5 ppm |
A1: 25 ppm | ± 10% of reading or ± 5% of full scale (*) |
A2: 50 ppm | ||||
Chlorine (Cl2)* | 0 – 3 ppm
0 – 10 ppm |
0.05 ppm
0.1 ppm |
A1: 0.5 ppm | |
A2: 1.0 ppm | ||||
Hydrogen Cyanide (HCN) | 0 – 15 ppm | 0.1 ppm | A1: 5 ppm | |
A2: 10 ppm | ||||
Nitrogen Dioxide (NO2) | 0 – 20 ppm | 0.05 ppm | A1: 3 ppm | |
A2: 6 ppm | ||||
Sulfur Dioxide (SO2) | 0 – 100 ppm | A1: 2 ppm | ||
A2: 5 ppm | ||||
IR Sensors | ||||
Carbon Dioxide (CO2)
|
0 – 10% vol | 25 ppm 0.2% | A1: 5,000 ppm | ± 5% of reading or ± 2% of full scale (*) |
TWA 5,000 ppm | ||||
STEL 10,000 | ||||
Methane (CH4)* | 0 – 100% LEL/
0 – 100% Vol. |
Vol. 1% LEL / 0.5% Vol |
A1: 10% LEL | |
Hydrocarbons** | 0 – 100% LEL/ 0 – 30% Vol. |
A2: 50% LEL | ||
PID Sensors | ||||
VOC | 0 – 50,000 ppb | 1 ppb (< 5000 ppb) 10 ppb (> 5,000 ppb) |
A1: 5,000 ppb | +10% of reading + 1 digit |
A2: 10,000 ppb | ||||
0 – 6,000 ppm | 0.1 ppm (< 600 ppm) 1 ppm (> 600 ppm) |
A1: 400 ppm | ||
A2: 1,000 ppm |
GX-6000 Webinar
SDM-6000
Sampling Hose & Probe
Red Boot
- Overview
-
Overview of the Benzene Personal PID gas monitor:
The GX-6000 personal Benzene PID gas monitor is the lightest and smallest Benzene monitor created, weighing in at only 600 grams with the rubber boot.
Simply put, the GX-6000 is a game changer with Benzene Specific measuring mode. With a library of over 300 VOC’s and the addition of the highly selective 10.0 eV PID sensor, the GX-6000 can now detect the BTEX (benzene, toluene, ethylbenzene, and xylene) family of VOC’s up to 100 ppm. By using filtering tubes and the tube holder, the GX-6000 can monitor specifically for Benzene at a range of 0 – 50ppm (with resolutions as small as 0.01) by filtering out all other VOC’s. The filter tube holder has a unique break away design for accidental damage. The filter tube also has a safe and convenient glass breaker for the filter tube tips.
The GX-6000 is the only PID instrument in the market with the ability to utilise two PID sensors in one instrument. When using the broad range 10.6eV PID for preliminary investigation, it is possible to switch to benzene mode for a highly accurate measurement.
The GX-6000 has a rugged design built for the nastiest environments. It is equipped with a removable impact-resistant rubber boot and a dust and water resistant enclosure with an IP-67 rating. With 5 bright LED lights on 3 sides of the instrument, alarms are easily seen from a variety of perspectives. With vibration and loud audible alarm, GX-6000 will easily alert users when needed.
- Specifications
-
Specifications:
Alarm Types Gas alarm Man-down Trouble Low and high adjustable 1st – self reset 2nd – latching
Low flow Sensor connection
Low battery
Circuit error
Calibration range
Sample Method Sample draw pump, flow rate 0.45 LPM, 30m sample range Display Digital LCD, autobacklite, peak bar display, auto display rotation, and customize order of gases Audible Alarm 95 dB at 1 ft. Languages English, German, Japanese, Korean, Russian Continuous
Operation *14 hours on Lithium-ion battery pack. Complete recharge in 3 hours. 8 hours Alkaline battery pack (3 AA size within each pack)
Operating Temp.
& Humidity-4°F to 122°F (-20°C ~ +50?), 0- 95% RH non condensing Dimensions / Weight 2.6” W x 7.8” H x 2” D (H200 x W68 x D52 mm)
14 ounce (400g) (with lithium battery pack)Case Material Dust and water resistant with IP-67 approval (exclude sample hose and probe) Safety Design / Approvals
Explosion proof: IECEx , ATEX Standard
Features- 2 Smart sensors slots auto recognise sensor changes
- Library of over 600 VOC’s
- User defined VOC list. Up to 30 frequently used
- Recent VOC list for last 8 selected gases
- Pump and circuit status indicators
- User & station ID selection menu
- Datalogging (interval, alarm trend, station and user ID)
- Snap logging – on demand data logging
- Demand zero / Auto zero
- BUMP and CAL expiration alarm
- Flashlight
- IrDA communication
- LEL sensor protection mode
Standard
Accessories- Rubber boot
- Belt clip
- Hand strap
- Tapered nozzle
- Probe with hydrphobic filter and 3’ teflon lined hose
Optional
Accessories- Lithium-ion battery pack & charger / alkaline battery charger
- Filter set • IrDA cable
Warranty Two years material (including sensors) and workmanship. One year for PID sensor - Detectable Ranges
-
Gas Measuring Range Minimum Display Alarm Set Points Accuracy
* Which ever is greaterPID Sensors 10.6 PID Lamp
600
VOC’s0 – 50,000 ppb 1 ppb
(< 5000 ppb)
10 ppb
(> 5,000 ppb)A1: 5,000 ppb +5% of full scale A2: 10,000 ppb 0 – 6,000 ppm 0.1 ppm
(< 600 ppm)
1 ppm
(> 600 ppm)A1: 400 ppm A2: 1,000 ppm 10.0 PID Lamp Benzene Specific No Alarms
300 VOC’s 0 – 100 ppmBenzene Specific
0.1 – 50 ppm0.01 ppm (0 ~ 10 ppm)
0.1 ppm (10 ~ 50 ppm)no alarms 300 VOC’s
0 – 100 ppm0.01 ppm (0 ~ 10 ppm)
0.1 ppm (10 ~ 100 ppm)A1: 5 ppm A2: 10 ppm Catalytic Hydrocarbons (CH4, std) 0 – 100% LEL 1% LEL A1: 10%LEL ± 5% of
reading or ± 2% LEL (*)A2: 50%LEL Galvanic Oxygen (O2) 0 – 40% Vol. 0.1% Vol. A1: 19.5 Vol. ± 0.5% O2 A2: 23.5 Vol. Electro Chemical Carbon Monoxide (CO) 0 – 500 ppm 1 ppm A1: 25 ppm ± 5% of reading or ± 5 ppm A2: 50 ppm TWA 25 ppm STEL 200 ppm Hydrogen Sulfide (H2S) 0 – 100 ppm 0.5 ppm A1: 5.0 ppm ± 5% of reading or± 2 ppm A2: 30.0 ppm TWA 1.0 ppm STEL 5.0 ppm Toxics (EC Sensors) Ammonia (NH3)* 0 – 75 ppm 0 – 400 ppm
0.5 ppm 5 ppm
A1: 25 ppm ± 10% of reading or ± 5% of full scale (*) A2: 50 ppm Chlorine (Cl2)* 0 – 3 ppm 0 – 10 ppm
0.05 ppm 0.1 ppm
A1: 0.5 ppm A2: 1.0 ppm Hydrogen Cyanide (HCN) 0 – 15 ppm 0.1 ppm A1: 5 ppm A2: 10 ppm Nitrogen Dioxide (NO2) 0 – 20 ppm 0.05 ppm A1: 3 ppm A2: 6 ppm Sulfur Dioxide (SO2) 0 – 100 ppm A1: 2 ppm A2: 5 ppm IR Sensors Carbon Dioxide (CO2) 0 – 10% vol 25 ppm 0.2% A1: 5,000 ppm ± 5% of reading or ± 2% of full scale (*) TWA 5,000 ppm STEL 10,000 Methane (CH4)* 0 – 100% LEL/ 0 – 100% Vol.
Vol. 1% LEL /
0.5% VolA1: 10% LEL Hydrocarbons** 0 – 100% LEL/
0 – 30% Vol.A2: 50% LEL PID Sensors VOC 0 – 50,000 ppb 1 ppb
(< 5000 ppb)
10 ppb
(> 5,000 ppb)A1: 5,000 ppb +10% of reading + 1 digit A2: 10,000 ppb 0 – 6,000 ppm 0.1 ppm
(< 600 ppm)
1 ppm
(> 600 ppm)A1: 400 ppm A2: 1,000 ppm - Downloads
-
- Videos
-
GX-6000 Webinar
- Accessories
-
SDM-6000
Sampling Hose & Probe
Red Boot