AC-3 Photoelectron Spectrophotometer

AC-3 Photoelectron Spectroscopy in Air (PESA)

Features

  • Atmospheric pressure operation (unique in the world)
  • Estimate work function, ionisation potential, density of states (DOS)
  • Measure the thickness of thin films on the material surface (less than 20 nanometers)
  • Measure the sample of powder and liquid
  • Low photo-excitation energy (4.0 – 7.0eV)
  • Measurement for just 5 minutes
  • Easy sample introduction and removal
  • Full computer control
  • Easy to measure the Density of States (DOS) around the highest occupied orbital, work function and ionisation potential
  • Information for tiny surface on nanometer order and tiny thin film thickness (0-20nm) can be measured
  • Powder and liquid samples that are unable to be brought into the vacuum can also be measured
  • Measurement taken in just 5 minutes (when measured in 5.0 – 7.0eV energy search range with step 0.1eV)

Applications

  • Measurement of ionisation potential on organic materials used in solar cells, fuel cells, organic EL and organic TFT
  • Measurement of ionisation potential on the photocatalyst
  • Evaluation of MgO film quality on plasma display panel (PDP)
  • Measurement of electronic states for carbon nanotube and fullerene
  • Measurement of electronic states on carbon thin film and diamond thin film
  • Measurement of electronic states for cosmetics, medicine and chemicals
  • Measurement of ionisation potential on sensitive materials
  • Measurement of contamination and oxidised film thickness for electrode, lead frame, contacts, steel plate, silicon wafer and compound semiconductor wafer
  • Measurement of work function on metallic materials for various electrodes

AC-3 Photoelectron Spectrophotometer Overview

The newly developed AC-3 is a photoelectron spectrophotometer operating at atmospheric pressure, using an open counter as the electron detector. The unit has been marketed mainly for the study of organo-electronic materials for organic EL and photocopying.

Additional Benefits

  • Easy to measure the Density of States (DOS) around the highest occupied orbital, work function and ionisation potential.
  • Information for tiny surface on nanometer order and tiny thin film thickness (0-20nm) can be measured
  • Powder and liquid samples that are unable to be brought into the vacuum can also be measured
  • Measurement taken in just 5 minutes (when measured in 5.0 – 7.0eV energy search range with step 0.1eV)

AC-3 Photoelectron Spectrophotometer Specifications

Model AC-3
Measuring Principle Low-energy electron counter method
Electron Detector Open counter
Ultraviolet Lamp Deuterium lamp with lamp house
Spectrometer Nitrogen substitution type grating method monochrometer
Energy Search Range 4.0 – 7.0eV (310 – 177nm)
Operating Condition 15 – 35°C, below 60% RH
Repeatability Precision

(Standard Deviation)

Work function 0.02eV
Measuring Time Approx 5 minutes (when measuring in 5.0 – 7.0eV energy scanning range with step

0.1eV)

Radiating Ultraviolet Rays Spot size: 2 x 5mm (slit 1.00mm)

Maximum intensity: above 100nW (at 5.9eV)

Sample Size 30 x 30mm max (max thickness 10mm), 1 sample measurement
Utility Required Power source: 100 VAC, 50/60 Hz, 5A (max)

Compressed Air: 0.5 – 0.6MPa, 5L/min

Nitrogen Gas: 0.1MPa, 2L/min (while measured), 5L/min (while purging)

Software AC-3 for Windows (functions to measure photoelectron spectrum, radiation light

spectrum, threshold energy etc)

Approvals CE
Overall Dimensions Approx 740 (w) x 1080 (h) x 680 (d) mm (with casters)
Weight Approx 120 kg
Overview

AC-3 Photoelectron Spectrophotometer Overview

The newly developed AC-3 is a photoelectron spectrophotometer operating at atmospheric pressure, using an open counter as the electron detector. The unit has been marketed mainly for the study of organo-electronic materials for organic EL and photocopying.

Additional Benefits

  • Easy to measure the Density of States (DOS) around the highest occupied orbital, work function and ionisation potential.
  • Information for tiny surface on nanometer order and tiny thin film thickness (0-20nm) can be measured
  • Powder and liquid samples that are unable to be brought into the vacuum can also be measured
  • Measurement taken in just 5 minutes (when measured in 5.0 – 7.0eV energy search range with step 0.1eV)

Specifications

AC-3 Photoelectron Spectrophotometer Specifications

Model AC-3
Measuring Principle Low-energy electron counter method
Electron Detector Open counter
Ultraviolet Lamp Deuterium lamp with lamp house
Spectrometer Nitrogen substitution type grating method monochrometer
Energy Search Range 4.0 – 7.0eV (310 – 177nm)
Operating Condition 15 – 35°C, below 60% RH
Repeatability Precision

(Standard Deviation)

Work function 0.02eV
Measuring Time Approx 5 minutes (when measuring in 5.0 – 7.0eV energy scanning range with step

0.1eV)

Radiating Ultraviolet Rays Spot size: 2 x 5mm (slit 1.00mm)

Maximum intensity: above 100nW (at 5.9eV)

Sample Size 30 x 30mm max (max thickness 10mm), 1 sample measurement
Utility Required Power source: 100 VAC, 50/60 Hz, 5A (max)

Compressed Air: 0.5 – 0.6MPa, 5L/min

Nitrogen Gas: 0.1MPa, 2L/min (while measured), 5L/min (while purging)

Software AC-3 for Windows (functions to measure photoelectron spectrum, radiation light

spectrum, threshold energy etc)

Approvals CE
Overall Dimensions Approx 740 (w) x 1080 (h) x 680 (d) mm (with casters)
Weight Approx 120 kg